Echo 525 Acoustic Liquid Handler
For aqueous-based acoustic transfers
- Designed for biochemical and genomic sample and reagent transfer
- Improve data quality while saving on reagent costs
- Reagent reservoir for larger volume reagent transfers
- Read more about Echo Acoustic Technology here
Explore Echo 525 Models
Liquid Handling Reimagined
Introducing the Echo 525 Liquid Handler, the first rapid acoustic liquid handler designed specifically for biochemical and genomics reagent transfer. The Echo 525 Liquid Handler and the Echo Software Applications enable scientists to spend more time looking at data. Protocol creation and implementation in the Echo Software Applications are fast and does not require extensive training. Designed for use with the Echo System, the Access Laboratory Workstation expands the Echo 525 Liquid Handler's utility and further frees up time.
Echo 525 Liquid Handler Features
Key Benefits
- Highly precise and accurate low volume transfers enabling assay miniaturization
- Builds assays 50–100 times faster than traditional liquid handlers giving the lab higher capacity from one instrument
- Suite of Echo Software Applications guide creation of complex protocols for a variety of applications
- Eliminates pipetting variances that affect assay success
- Integrates into the Labcyte Access Laboratory Workstation for increased throughput
Build Complex Assays Quickly
The flexibility, speed, accuracy, and precision of the Echo 525 Liquid Handler enable researchers to build assays that are not possible with traditional liquid handlers. The use of acoustic sound energy removes the traditional boundaries of tip and pin tool based liquid handlers. Transfer variable fluid heights and viscosities from one source plate to enable rapid assay creation in one fluid transfer run. Say goodbye to fixed plate designs as the Echo Liquid Handler moves fluids from any well to any well enabling researchers to design the experimental plates they need to make new discoveries.
Flexible and Reliable Transfer of Genomic Reagents
The Echo 525 Liquid Handler transfers a wide range of fluids during one transfer including distilled water, buffers, nucleic acids, and reagents containing up to 50% glycerol. With its versatile fluid range, transfer volume range, any well to any well transfer capability, and powerful applications software the Echo 525 Liquid Handler supports a wide range of genomic assay workflows:
- Gene Synthesis
- PCR and qPCR assay set-up
- RNA and DNA NGS sequencing workflows
- Sanger sequencing assay set-up
- Single-cell sample preparation
Echo 525 Acoustic Liquid Handler Specifications
Drop volume | 25 nL |
Volume transfer range | 25 nL – 5 µL one well to one well |
Transfer accuracy | <10% deviation from target volume |
Transfer precision | <8% CV |
Reagents supported | PCR and qPCR reagents, synthetic biology reagents, NGS reagents, primers and probes, cell culture media, proteins, nucleic acids, up to 50% glycerol, serum, plasma, and antibodies |
Source labware compatibility | EchoQualified 384 Polypropylene, 384 Polypropylene Plus, Reservoir, Microplates |
Destination labware compatibility | All Echo Qualified Microplates and most ANSI-compliant/SBS-standard microplates in 96-, 384-well formats 8 -16 mm in height. Use of 1536-well plates as destination plates is application dependent |
Dimensions | 53.9 cm width x 68.3 cm depth x 92.5 cm height (21.2 x 26.9 x 36.4 inches) |
Weight | 128 kg (283 lbs) |
Working envelope | Additional 25.4 cm (10 inches) on top, 2.5 cm (1 inch) on sides, 7.6 cm (3 inches) at front, 30.5 cm (12 inches) at back |
Electrical | AC 120V, 50/60 Hz, 10 A or AC 230V, 50/60Hz, 5 A |
Operating temperature | 21°C ± 5°C (70°F ± 9°F) |
Vacuum | House supply minimum 200 Torr (266 mbar, 22 inches Hg Vac) OR Standalone vacuum with surge tank, minimum pump speed 2.0 m3/hr (70.68 cu ft/hr), minimum pressure 100 Torr (133 mbar, 26 inches Hg Vac) |
Fluid class calibration | No calibration needed |