ISO 11171: Particle Counter Performance and Calibration
ISO 11171:2016 Calibration Schedule
Instrument Status |
Recommended Steps |
Routine calibration |
Sizing calibration required (Clause 6) |
Suspect calibration (> 6-12 months) |
Sizing calibration required (Clause 6) |
Repaired/adjusted sensor |
Entire calibration must be performed – Sizing Calibration, Preliminary Check, Coincidence Error & Flow Rate Limits, Resolution & Accuracy required |
Repaired/adjusted volume measurement components |
Preliminary Check required |
Repaired/adjusted optics |
Sizing calibration, Preliminary Check, Coincidence Error Limits, Resolution, Accuracy required |
New LPC or no prior 1171 calibration |
Entire calibration to be performed – Sizing Calibration, Preliminary Check, Coincidence Error & Flow Rate Limits, Resolution & Accuracy |
Sensor cleaned |
No action necessary |
Cables or peripheral equipment attached |
No action necessary |
Plumbing lines/connections replaced |
No action necessary |
Operation performed, no instrument disassembly, sensor or volume measurement system involved |
No action necessary |
NOTE: Annexes A, B, C and D can be performed by an individual lab or by the manufacturer prior to equipment delivery. Although full ISO 11171 calibration is time consuming, subsequent routine sizing calibrations can be done significantly quicker.