Do EU GMP or cGMP guidelines require the use of a specific particle counter for air particle cleanliness classification and monitoring?
EU GMP and cGMP point to ISO 14644-1 for the methodology to determine cleanroom air particle cleanliness classification. ISO 14644-1 states that ISO 14644-2 should be used for routine environmental monitoring. Both ISO 14644-1 and -2 mandate that particle counters must be calibrated to ISO 21501-4.